Image of Simultaneous determination of 11 related impurities in propofol by gas chromatography/tandem mass spectrometry coupled with pulsed splitless injection
technique

LITERATURE

Simultaneous determination of 11 related impurities in propofol by gas chromatography/tandem mass spectrometry coupled with pulsed splitless injection technique


A variety of related impurities, including starting materials, process impurities, and degradation products, can be detected in propofol. In this article, a sensitive and selective GC-MS/MS method using pulsed splitless injection technique for the determination of 11 main related impurities in propofol in one chromatogram is investigated. This method is extensively validated for its linearity, recovery, precision, LOD, and LOQ, and is able to detect trace-level
related impurities (LOD = 0.2–5.6 g/g) in propofol bulk drug. Stressed tests proposed that oxidative degradation, photolytic degradation, and heat are the main causes for the formation of degradation products in propofol.


Availability

Recofol-003Available

Detail Information

Series Title
-
Call Number
-
Publisher : .,
Collation
-
Language
English
ISBN/ISSN
-
Classification
NONE
Content Type
-
Media Type
-
Carrier Type
-
Edition
-
Subject(s)
Specific Detail Info
J. Sep. Sci. 2013, 36, 1959–1966